|
Possible Cause |
Prevention/Solution |
|
Decreasing
Retention Times |
|
|
Loss of bonded stationary phase |
Replace column Operate at pH 2-8 for silica-based RP columns |
|
Active groups on a stationary phase |
Use organic modifier in the mobile phase Increase buffer strength |
|
Increasing flow rate |
Check and adjust the pump flow rate |
|
Column overloaded |
Reduce the amount of sample injected Use a column with a larger i.d. |
|
Increasing
Retention Times |
|
|
Changing mobile phase composition |
Cover solvent reservoirs Prepare fresh mobile phase |
|
Loss of
bonded stationary phase |
Replace
column |
|
Decreasing flow rate |
Check and adjust the pump flow rate Check for leaks in the system, including pump
seals |
|
Bubbles in mobile phase |
Check flow rate and pressure Degas mobile phase |
|
Fluctuating Retention Times |
|
|
Insufficient column equilibration |
Equilibrate column longer between runs Condition the column with a concentrated sample |
|
Change in mobile phase composition |
Check make-up of the mobile phase and make-up new
if necessary Check proportioning-valve accuracy |
|
Insufficient buffer capacity |
Use buffer concentrations >20mM |
|
Fluctuating column temperature |
Stabilize ambient temperature Thermostat the column |
0 Comments